Laboratory Test Request

IEC Medical Device Design Verification

Project Description

USA, CANADA OR EUROPE LABORATORY
Large Corporation needs Medical Device Laboratory needed for Medical Device Design Verification to IEC 60601 including the following:

Place a medical device inside a carrying accessory on a mannequin . Measure the sound, SPL, of an alarm from the medical device to be above 25 dB at 1 m in all six orthogonal positions (front, back, top, bottom, and both sides). Note that either the sound receiver or the mannequin can be moved to reach the six positions. For information, we would also like the SPL from a simulated head position of the carrying accessory on the mannequin.

Project Information

Number:21-00670

Industry:Medical Devices